![Thermo Scientific Heracell Vios 160i CR Operating Instructions Manual Download Page 161](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i-cr/heracell-vios-160i-cr_operating-instructions-manual_1104558161.webp)
9
Maintenance
Replacing the HEPA Filter in the Workspace
Thermo Scientific
Heracell Vios 160i / 250i CR
9-7
Figure 9-5.
CO
2
Calibration Procedure
4. Enter the measured value (target value):
The targeting value can be increased or reduced in increments; if you keep the + key or the – key
depressed, the function switches to a rapid increase/reduction; after approx. 3 seconds, another
increase/reduction occurs.
To increase the targeting value:
• Press the + key.
To reduce the set value:
• Press the - key.
5. Accept and save the target value:
• Press the
Enter
key.
• Press the
Save
key.
• The system returns to the main menu. The CO
2
display shows the actual value currently measured
in the workspace.
Replacing the HEPA Filter in the Workspace
The HEPA filter in the workspace is installed under a plastic airbox at the bottom of the workspace,
right before the rear wall.
Steps for replacing the HEPA filter:
1. Power down the unit, shut off the gas supply and allow the workspace to vent.
2. Remove the center and bottom shelves out of the workspace.
Note
Excessive CO
2
content:
Excessive CO
2
content after the calibration can be reduced by leaving the device doors open for
approx. 30 seconds.
Value reset:
If the value isn’t changed within the next 30 seconds, the system automatically exits the menu, and
the most recently confirmed value is preserved.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...