2
Installation of the Device
Stacking
Thermo Scientific
Heracell Vios 160i / 250i CR
2-7
Stacking Variants
For details see assembly instruction of stacking adapter kit.
Installing the Support Stand with Castors
Figure 2-7.
Support Stand with Castors without Supporting Feet (160i Unit)
• Position unit with device feet [1] matching all 4 staple pads [2].
For positioning, do not lift the device by using the doors or components attached to the device
(e.g. control box on rear panel) as lift points.
• Secure the support stand by unscrewing the 4 bolts [3].
Possible stacking combinations
Bottom stacking adapter
Heracell Vios 160i CR
Steri-Cycle i160 CR
Heracell Vios 250i CR
Steri-Cycle i250 CR
T
op stacking adapter
Heracell Vios 160i CR
Steri-Cycle i160 CR
50148171
Heracell Vios 250i CR
Steri-Cycle i250 CR
50154522
CAUTION
Providing Stability
Make sure that the support stand castors are locked during operation of the device and that the
castors are facing forward.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...