8
Cleaning and Disinfection
Steri-Run Sterilization Routine
Thermo Scientific
Heracell Vios 160i / 250i CR
8-11
• remaining run time.
Figure 8-6.
Activate Steri-Run
Canceling Steri-Run
The Steri-Run sterilization routine can be interrupted at any time.
1. Cancel Steri-Run:
• Press the
Stop
key.
• If the
Stop
key in the status display is depressed, the Steri-Run stop dialog box is displayed for
a safety scan. The routine can now be permanently canceled or resumed.
2. To stop Steri-Run:
• Press the
End
key.
• The error message is displayed.
• If the error message is confirmed, the system returns to the main menu.
3. To resume Steri-Run:
• Press the
Back
key.
• The system returns to the status display, and the sterilization routine is resumed.
4. To interrupt Steri-Run from the status display:
• Press the
Stop
key.
• As a safety scan, the Steri-Run - Stop dialog window appears. Proceed with work step 2.
Note
As long as the workspace temperature is above 65°C, the door remains locked! For manual
unlocking, see Chapter 8-13.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...