8
Cleaning and Disinfection
Steri-Run Sterilization Routine
8-14
Heracell Vios 160i / 250i CR
Thermo Scientific
Opening the Door after Canceling Steri-Run
The door cannot be opened immediately after the Steri-Run sterilization routine has been aborted by
the user or due to an error.
To release the outer door lock before the temperature has dropped to a non-hazardous level of 65°C,
you must use the emergency release lever (item 2 in
):
• Locate the emergency release lever (2) at the bottom of the incubator. It is located on the left side
of the device, approx. 5 centimeters inwards from the main power switch (4), as shown in
below).
Figure 8-9.
Door Lock Latch and Emergency Release on Underside of Incubator
• Reach under the bottom panel and remove the round cover (3/Fig. 8-9).
• Insert a finger through the hole and pull the emergency release lever above it towards the power
switch.
• The door lock disengages and releases the outer door for opening
• Replace the round cover (3/Fig. 8-9).
CAUTION
Hot surfaces!
The handle and the glass door, the interior panel of the outer door as well as the surfaces of the shelf
system and of the workspace become extremely hot during the Steri-Run sterilization routine.
During the routine run or immediately after completion of the run, always wear safety gloves when
touching these surfaces!
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...