7
Installation
Installation of GenPure xCAD Plus system, wall version
Thermo Scientific
GenPure
39
12
Mount the feedwater connecting kit together and
connect it to the feedwater inlet line. Connect the
other end of the hose to the feedwater connector
of the system by unscrewing the fitting. After this
put the hose through the fitting and mount the
white O-ring on it. Screw the fitting back to the
system.
Only feedwater that has been pretreated by
reverse osmosis, ion exchange or distillation is to
be used.
13
Connect the 0.31” (8 mm) o.d hose to the rinse
water connection of the system (see step 11) and
make a gravity fall (pressureless) connection from
the system to the floor drain. The drain to the
sewer must be max. are 1m (1.09 yards) above the
rinsing water connector of the unit.
14
Screw the final filter in counter clockwise
direction (see red arrow in the picture) into the
1/4" female thread of the xCAD dispensing valve.
Step
Action
Figure
Feedwater
connecting kit
Fitting
Feedwater
connector
unit
Hose 0.31”
8 mm
o.d
White
O-ring
Rinse water
connection
White
O-ring
Hose 0.31”
8 mm o.d
Fittin
g
1/4" female thread
connection
Final filter 0.2
μ
m
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...