12
System control
Operating modes
Thermo Scientific
GenPure
67
Water dispensing via volumetric control
Ultrapure water systems which are equipped with the volumetric dispense option can dispense a preset
volume of water.
As soon as the Nonstop-mode is selected, a liter volume is shown in line 3 of the display. This is the
volume of ultrapure water that was last dispensed.
A single press on the Enter-button enables this volume value to be changed within the range from 0.01
to 65.5 liters by means of the arrow-buttons. The UV-button can be used to position the cursor at the
particular digit that you wish to change.
A second press on the Enter-button causes the volume of water that has been set to be dispensed. The
liter volume shown in the display is the actual volume dispensed. Dispensing stops as soon as the set
volume is reached.
Dispensing can be stopped at any time by a further press on the Enter-button. This enables small
volumes to be dispensed by two successive presses on the Enter-button. One press starts dispensing
and, when the wanted amount has been dispensed, a second press stops dispensing. The button on the
dispenser has the same function as the Enter-button.
Volumetric dispense is supported in all versions.
The display shows:
OFF mode
A second press on the ON/Off-button causes the display to go dark and all text output on the display
to be extinguished. No outputs are now switched.
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...