13
Maintenance
Change the ultrapure cartridge
86
GenPure
Thermo Scientific
7
Open the supply of feedwater and switch the
system on again.
8
The Code to do this transaction please refer from
the Code table „code lock” found in chapter
“Code lock”
on
page 82
. You need a level one
code.
a.
Go in the Menu to the point “change
filter cartridge” and press enter.
b. Enter new serial number of the ultrapure
cartridge in by pushing the button
nonstop or Interval to change the digits
and the UV button to go to the next
value.
c.
When you are finished, press enter and
the new serial number is saved. You can
only use a serial number one time.
9
Discard at least 5 liters of water.
Step
Action
Figure
Feedwater
supply
a.
b.
c.
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...