Safety precautions
• Do not start to install and operate the system until you have read through the corresponding
information given in these operating instructions.
• Lifting and carrying the ultrapure water system, e.g. to the installation location, should be carried
out by two people. To lift it, each person takes hold of it under the base plate at two corners.
• The CE-mark is invalidated if constructional changes are made to the system, or if products of
other manufacturers are installed in it.
• Protect the system from frost. The temperature in the area in which the system is installed is not to
go below +2°C or above +40°C.
• Observe all appropriate rules and regulations, including the valid accident prevention regulations,
which are applicable at the location where the system is installed.
• The feedwater pressure must be at least 0.1 bar and at max. 6 bar or 1.45 to 87 PSI. When the
feedwater pressure is higher, install an external pressure reducer.
• A low pressure check valve is recommended to prevent back flow of feedwater from water system.
• A grounded 100-250V, 50/60Hz socket must be available (see
“Technical specifications”
on
page 19
).
• Access to the power supply cord and plug may never be restricted or obstructed.
• The installation area must have a drain at floor level with at least a nominal out diameter of 63mm
or 2.481 inch (DN 50 pipe). Should no such drain be available it is recommend to install a water
watcher (only for European specification). Otherwise the manufacturer will not accept liability for
any possible water damage.
Observe these safety precautions for your own safety.
Thermo Scientific Barnstead Ultrapure Water Systems are modern water purification
systems intended solely for the treatment of potable water or water of ASTM Type II
quality. The water it produces is not fit for drinking.
Work may only be performed on the unit electronics when the unit has been
switched off and when ESD protection is in place. Only specially trained personnel
may work on the unit‘s electronics.
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...