13
Maintenance
Disinfection
Thermo Scientific
GenPure
89
5
Push the menu button until “Enter code” is
displayed
The Code to do this transaction please refer from
the Code table under chapter
“Code lock”
on
page 82
. You need a level 1 code.
a.
Select “Disinfection” from the system
menu and press “Enter”.
b. Confirm the Disinfection Cartridge has
been loaded by pushing “Enter” again
c.
The disinfection process will begin.
The disinfection program is finished after approx
30 min and is adjustable in the OEM Menu.
6
Switch the system off and shut off the water
supply.
See step 1.
7
Remove the disinfectant cartridge, empty and dry
it and put in the yellow stoppers that you have
saved for later use. Save the disinfection cartridge
for later use.
See step 5 under chapter
“Change the ultrapure
cartridge”
on
page 85
.
8
Before dispensing water from the system, let
water run out for approx 15 minutes.
The system is then ready for use.
Step
Action
Figure
a.
b.
c.
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...