13
Maintenance
Maintenance intervals
84
GenPure
Thermo Scientific
Regular servicing of your system ensures that the quality of water is maintained. We recommend a
service contract with a factory authorized service company to ensure that the system is properly
maintained. You then have the certainty of a high operational, safe, and reliable water purification
system.
To ensure error-free operation, your system must be checked, serviced and cared for at regular time
intervals in accordance with these operating instructions. For this reason, the operating instructions
must be readily available to operating and maintenance staff at all times, and be carefully followed.
Calibration of the conductivity is only to be carried out and recorded by a factory-authorized service
technician.
Cleaning and disinfection should be performed at least once yearly, or when the ultrapure cartridge is
replaced, or when bacteria is present in the product water.
Maintenance intervals
Consumable materials are to be replaced according to the directions below. The intervals were
determined for the average user and are completely dependent on the actual feed water quality and
volume of water used daily.
*Please keep in mind that the life of your consumable is directly dependent on the quality of the feed
water and the amount of water used daily.
Control and maintenance work on electrical systems are only to be carried out by an
appropriately trained, skilled electrician.
Unplug the system from the power outlet for all Maintenance work on the system.
Material
Flow chart no.
Catalog no.
Interval
Other problems
Ultrapure
cartridge
F1
09.2005
up to 12
Months
Or when the ultrapure water limiting
value is exceeded, whichever is shorter.
Longer usage can result in bacterial growth
on the resin.
Sterile 0.2 micron
filter
F2
09.1003
up to 12
Months
Or flow rate is noticeably slower.
Ultrafiltration
membrane (only
applicable for
systems with a UF
filter)
F3
50133980
up to 24
Months
Or if there is endotoxin breakthrough in
product water or when the water flow rate
is markedly slower.
UV-lamp (only
applicable for
systems with a UV
lamp)
UV1
09.2002
up to 24
Months
Or unless system indicates the lamp needs
to be replaced.
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...