7
Installation
Installation of GenPure xCAD Plus system, bench version
Thermo Scientific
GenPure
31
Installation of GenPure xCAD Plus system, bench version
• Control after installation all tubings that the tubing have the right positions on
the systems panel and there is no any leakage after open the feedwater supply tap.
• To avoid a book trip over the tubings, observe that the tubings are not lay over the floor.
Step
Action
Figure
1
Either place the system on the intended surface or
hang it on a wall. For wall mounting using the
included wall mounting hardware.
See under chapter
“Wall mounting GenPure
xCAD Plus system”
on
page 46”
.
2
Release the cartridge cover by pressing the push
button.
3
Locate the ultrapure cartridge and fit the
cartridge into the system.
4
Push each of the quick connectors onto the
cartridge. You will know they are attached when
an audible “click” is heard.
Fit the cartridge cover on again.
5
Mount the feedwater connecting kit together and
connect it to the feedwater inlet line. Connect the
other end of the hose to the feedwater connector
of the system by unscrew the fitting. After this
put the hose through the fitting and mount the
white O-ring on it. Screw the fitting back to the
system.
Only feedwater that has been pretreated by
reverse osmosis, ion exchange or distillation is to
be used.
Push-button
Cartridge cover
Inlet
Outlet
Ultrapure
cartridge
Quick
connectors
Feedwater connecting kit
White O-ring
Hose 8 mm o.d
Feedwater
connector
unit
Fitting
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...