12
System control
Measuring cell error recognition
82
GenPure
Thermo Scientific
Error message:
When a error message is shown in the display, e.g. for the ultrapure water limiting value, then the error
message is printed out on expiry of the sending interval.
Print-out:
Measuring cell error recognition
Minimum and maximum limiting values for each of the conductivity measuring cells and the
temperature sensor are fixed. Should measured values go below or above these respectively, then it must
be assumed that a cable break has occurred. The appropriate error message
“Measuring cell LF1”
,
“Measuring cell LF2”
,
“Measuring cell LF3”
or
“Measuring cell Temp”
is then issued in line 1.
When resistances are in a region below 50
:
or above 20MW, then a cable break or a short-circuit can
be assumed.
These basic settings cannot be changed in any menu.
Code lock
To prevent unauthorized access to system control settings, changes to these settings can only be carried
out when a correct code number has been entered and confirmed with Enter.
In deviation to existing programs, control release can be given at three levels. Only the menu is released
for changes at the first level. Both the menu and the OEM menu are released at the second level. All
menus are released at the third level.
Code numbers:
Each access via the code is printed out by the printer (RS 232) complete with date, time and the code
number used.
The display shows:
No.
Menu
No.
Menu + OEM menu
No.
All levels
1
0150
4
0450
7
0750
2
0250
5
0550
8
0850
3
0350
6
0650
9
0950
To prevent unauthorized access to system control settings, changes to these settings can only be carried
out when a correct code number has been entered and confirmed with Enter.
In deviation to existing programs, control release can be given at three levels. Only the menu is released
for changes at the first level (Code number 1, 2, 3). Both the menu and the OEM menu are released at
the second level (Code number 4, 5, 6). All menus are released at the third level (Code numbers 7, 8,
9).
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...