6
Connect the 8mm o.d hose to the rinse water
connection of the system (see step 5) and make a
gravity fall (pressureless) connection from the
system (connector 7) to the floor drain. The drain
to the sewer must be max. 1m (1.09 yards) above
the rinsing water connector of the Unit
7
a.
Connect the one end of the 0.31”
(8 mm) od hose to the xCAD backflow
connector on the system by unscrewing
the fitting. After this put the hose
through the fitting and mount the white
O-ring on it. Screw the fitting back to
the system.
b. The other end of the 0.31” (8 mm) o.d
hose you should connect to the xCAD
backflow connector of the xCAD Server.
Step
Action
Figure
White
O-ring
Fitting
Hose 8 mm
o.d
Rinsing water
connection
Connector GenPure
system
Fitting
Hose 0.31“
(8 mm) o.d
a)
Connector
xCAD
Server
Hose 0.31”
(8 mm) o.d
b)
White O-ring
Summary of Contents for Barnstead GenPure xCAD Plus System
Page 10: ...Contents 4 GenPure Thermo Scientific ...
Page 16: ...2 Safety precautions 10 GenPure Thermo Scientific ...
Page 22: ...3 Extent of delivery Available GenPure xCAD Plus versions 16 GenPure Thermo Scientific ...
Page 24: ...4 Intended Use 18 GenPure Thermo Scientific ...
Page 30: ...5 Technical specifications 24 GenPure Thermo Scientific ...
Page 32: ...6 The installation area 26 GenPure Thermo Scientific ...
Page 58: ...8 52 GenPure Thermo Scientific ...
Page 60: ...9 How the system functions 54 GenPure Thermo Scientific ...
Page 104: ...13 Maintenance Change and autoclave the Final filter 98 GenPure Thermo Scientific ...
Page 106: ...14 Waste disposal 100 GenPure Thermo Scientific ...
Page 110: ...15 Trouble shooting 104 GenPure Thermo Scientific ...
Page 114: ...16 Replacement parts GenPure 108 GenPure Thermo Scientific ...
Page 116: ...17 Consumable materials 110 GenPure Thermo Scientific ...
Page 118: ...18 Accessories 112 GenPure Thermo Scientific ...
Page 119: ...Thermo Scientific GenPure 113 19 Terminal assignments ...
Page 120: ...19 Terminal assignments 114 GenPure Thermo Scientific ...
Page 122: ...20 Maintenance records 116 GenPure Thermo Scientific ...