36
Query Syntax :STATus:QUEStionable:INSTrument:ISUMmary<n>:ENABle?
Parameter
<n>
1~3
<NR1>
0~32767
Return
parameter
<NR1>
0~32767
Example
:STAT:QUES:INST:ISUM1:ENAB 1
Sets the bit0 of the Questionable Instrument Summary Status
Enable register.
Query
example
:STAT:QUES:INST:ISUM1:ENAB?
>1
Bit0 is set.
4-9-14. :STATus:PRESet
Set
Description
This command resets the ENABle register, the PTRansistion filter
and NTRansistion filter on the Operation Status and Questionable
Status Registers. The registers/filters will be reset to a default
value.
Default Register/Filter Values
Setting
QUEStionable Status Enable
0x0000
QUEStionable Status Positive Transition
0x7FFF
QUEStionable Status Negative Transition
0x0000
QUEStionable Instrument Summary1 Status Enable
0x7FFF
QUEStionable Instrument Summary2 Status Enable
0x7FFF
QUEStionable Instrument Summary3 Status Enable
0x7FFF
Operation Status Enable
0x0000
Operation Status Positive Transition
0x7FFF
Operation Status Negative Transition
0x0000
Summary: The Questionable Status Enable registers and the
Operation Status Enable registers are both reset to 0.
The Questionable Status and Operation Status Positive
Transition filters are all set high (0x7FFF) and the Negative
Transition filters are all set low (0x0000).
I.e., only positive transitions will be recognized for the
Questionable Status and Operation Status registers.
Syntax
:STATus:PRESet
Example
:STAT:PRES
Sets the initial value to Questionable status and Operation status.
4-10. Source Commands
4-10-1. [:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For externally set
current levels (from the analog control connector) the set current
level is returned.