17
F-73 = 0~2
Parity 0 = none, 1 = odd, 2 = even
F-74 = 0 or 1
Stop bits:
0 = 1, 1 = 2
F-75 = 0
TCP: 0 = SCPI
F-76 = 0~30
UART address for multi-unit remote
connection.
F-77 = 0~3
Multi-Drop control:
0 = Disable, 1 = Master, 2 = Slave,
3 = Display Information
F-78 = 0~30
Multi-Drop status display
Displayed parameter: AA-S
AA: 0~30 (Address),
S: 0~1 (Off-line/On-line status).
3. The RMT indicator will turn on when a remote connection
has been established.
V
A
RMT
2-4-2. UART Function Check
Functionality
check
Invoke a terminal application such as Realterm.
To check the COM port No, see the Device Manager in the
PC.
Run this query command via the terminal application after
the instrument has been configured for either RS232 or
RS485 remote control.
*IDN?
This should return the Manufacturer, Model number, Serial
number, and Firmware version in the following format:
TEXIO,PFR-100L50,TW1234567,01.01.12345678
Manufacturer: TEXIO
Model name : PFR-100L50
Serial number : TW1234567
Firmware version : 01.01.12345678
^j (LF:Line Feed) can be used as the terminal character
when entering the queries/commands from a terminal
application.