47
Query
example
:SYST:CONF:OUTP:PON?
>0
Returns the power on output setting.
4-11-7. :SYSTem:CONFigure:OUTPut:EXTernal:MODE
Set
Query
Description
Sets the logic used to turn the output on or off when using an
external contact. This is the equivalent to the F-94 (External
Output Logic) power on configuration settings.
Syntax
:SYSTem:CONFigure:OUTPut:EXTernal:MODE {<NR1> | LOW |
HIGH}
Query Syntax :SYSTem:CONFigure:OUTPut:EXTernal:MODE?
Parameter
0 / HIGH Active High
1 / LOW Active Low
2
External control is not performed.
Return
parameter
<NR1>
Returns the logic setting.
Example
:SYST:CONF:OUTP:EXT:MODE HIGH
Sets the external logic mode.
Query
example
:SYST:CONF:OUTP:EXT:MODE?
>0
Returns the set value of external logic mode.
4-12. Communication system configuration Commands
4-12-1. :SYSTem:COMMunicate:ENABle
Set
Query
Description
Enables/Disables GPIB, USB or other remote interfaces such as
Sockets and the Web Server. (F-29)
This setting is only applied after the unit has been reset. Only one
interface can be enabled at the same time.
Syntax
:SYSTem:COMMunicate:ENABle {<Boolean> | OFF | ON, RS232
| RS485 | USBCDC | GPIB | SOCKets | WEB}
Query Syntax :SYSTem:COMMunicate:ENABle? {RS232 | RS485 | USBCDC |
GPIB | SOCKets | WEB}
Parameter 1
0 / OFF
Disables the selected interface.
1 / ON
Enables the selected interface.
Parameter 2
RS232
Select the RS232 interface
RS485
Select the RS485 interface
USBCDC
Select the USB-CDC interface
GPIB
Select the GP-IB interface
SOCKets
Select the Sockets interface
WEB
Select the Web Server interface
Return
parameter
<Boolean> Returns the status of the selected mode.