1. Configuration Settings
1-1. Setting Configuration Settings
Background
The normal configuration settings (F-01~F-61, F-71~F-78,
F-88, F-89) are used to configure or view system settings. Use
the following operation steps when configuring the interface
settings used in the Communication Interface chapter. (page 5)
Ensure the load is not connected.
Ensure the output is off.
Note
Configuration settings F-90~F-94 cannot be edited in the
Normal Function Settings. See the user manual for details.
F-20, F-21, F-25, F-30~F-35, F-78, F-89 settings can only be
viewed, not set.
Steps
1.
Press the Function key. The function key will
light up.
Function
M1
2.
The display will show F-01 on the top and
the configuration setting for F-01 on the
bottom.
3.
Rotate the Voltage knob to change the F
setting.
Voltage
Range
F-00~F-61, F-71~F-78,
F-88~F-94
4.
Use the Current knob to set the parameter for
the chosen F setting.
Current
5.
Press the Voltage knob to
save the configuration
setting.
ConF will be displayed
when successful.
Voltage
Exit
Press the Function key again to exit the
configuration settings.
The Function key light will turn off.
Function
M1