38
Parameter
OFF | 0
Disable the setting current limit.
ON | 1
Enable the setting current limit.
Return
parameter
<Boolean> Returns the setting in <boolean> format.
Example
:CURR:LIM:AUTO 0
Disables the current limit.
Query
example
:CURR:LIM:AUTO?
>0
Returns the limit state. Disables the current limit.
4-10-4. [:SOURce]:CURRent:PROTection:DELay
Set
Query
Description
Sets the Delay Time for OCP in seconds for turning the output off
(F-12). The delay is set to 0.
0
by default.
Syntax
[:SOURce]:CURRent:PROTection:DELay {<NR2> | MINimum |
MAXimum}
Query Syntax [:SOURce]:CURRent:PROTection:DELay?
Parameter
<NR2>
0.1~2.0 seconds, where 0 = no delay
MAXimum The maximum allowed delay time (2.0).
MINimum
The minimum allowed delay time (0.1).
Return
parameter
<NR2>
Returns the delay time in seconds
Example
:CURR:PROT:DEL MAX
Sets the OCP delay to the maximum (2.0).
Query
example
:CURR:PROT:DEL?
>2.0
Returns the current setting value. Set to 2.0 seconds.
4-10-5. [:SOURce]:CURRent:PROTection[:LEVel]
Set
Query
Description
Sets or queries the OCP level in amps.
Syntax
[:SOURce]:CURRent:PROTection[:LEVel] {<NRf>|MINimum |
MAXimum}
Query Syntax [:SOURce]:CURRent:PROTection[:LEVel]? [MINimum |
MAXimum]
Parameter
<NRf>
OCP level.
0%~110% of the rated current output in amps.
MINimum
Minimum OCP level.
MAXimum Maximum OCP level.
Return
parameter
<NR2>
Returns the OCP level.
Example
:CURR:PROT 5
Sets the OCP level.
Query
example 1
:CURR:PROT?
>+5.000
Returns the OCP level.