2-1-2. USB-CDC Remote Control Function Check
Please note that the port settings may not be confirmed if the
device driver is not used for this unit.
Please prepare the Terminal application (such as PuTTY or
RealTerm). Serial communication settings are as follows.
Baud rate
: 9600 bps
Data bits
: 8 bits
Parity bit
: None
Stop bits
: 1 bit
Flow control : None
Run this query command via the terminal after the instrument
has been configured for USB remote control.
*IDN?
This should return the Manufacturer, Model name, Serial
number, and Firmware version in the following format.
TEXIO,PFR-100L50,TW1234567,01.01.12345678
Manufacturer : TEXIO
Model name : PFR-100L50
Serial number : TW1234567
Firmware version : 01.01.12345678
Termination character of commands and queries use the
^j (LF: Line Feed).
2-2. GPIB Interface
2-2-1. GPIB Remote Interface
To use GPIB, you must select a model with a GPIB. This is a factory installed
option and cannot be installed the end-user. Only one GPIB address can be
used at a time.
Configure GPIB 1. Ensure the PFR-100 is off before proceeding.
2. Connect the GPIB cable (part number: GTL-258) from a
GPIB controller to the GPIB port on the PFR-100.
3. Turn the PFR-100 on.
4. Press the Function key to enter the Normal
configuration settings.
Function
M1
5. Set the following GPIB settings.
F-29 = 4
Enable the GPIB port
F-23 = 0~30
Set the GPIB address (0~30/ Default : 8)
6. Check to see the GPIB option is detected by the PFR-100.
The F-25 setting indicates the GPIB port status.
F-25 = 0
Indicates that the GPIB port is not detected.
F-25 = 1
Indicates that the GPIB port is available.