Theory of Operation
DSA8200 Digital Serial Analyzer and Modules Service Manual
2- 11
1
Clock recovery options are available (OC-192 and FEC10.709)
The 80C10 module is designed to test long wavelength (1310 and 1550 nm)
signals. This module is intended to be used as a test and measurement tool for
high bandwidth telecommunications with its high optical bandwidth >65 GHz.
1
Supported standards or data filtering rates include OC-768 and FEC43.02
(G.709).
1
30 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>30 GHz.
1
65 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>65 GHz.
1
There is no clock recovery option available.
The 80C10B module is designed to test long wavelength (1310 and 1550 nm)
signals. This module is intended to be used as a test and measurement tool for
high bandwidth telecommunications with its high optical bandwidth >80 GHz.
1
Supported standards or data filtering rates include OC-768, 40GBase--FR,
and FEC43.02 (G.709). Option F1 adds support for 100GBase--LR4+FEC,
100GBase--ER4+FEC, and 25.781 Gb/s (100GBase--LR4 and 100GBase--
ER4).
1
30 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>30 GHz.
1
65 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>65 GHz.
1
80 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>80 GHz. Not available with Option F1.
1
Option CRTP provides electrical signal outputs for use with the Tektronix
CR286A Option HS clock recovery device, or other compatible clock
recovery devices.
The 80C11 module is designed to test long wavelength (1100--1650 nm) signals.
This module is intended to be used as a test and measurement tool for high
bandwidth telecommunications with its high optical bandwidth >20 GHz.
80C10 Optical Sampling
Module
80C10B Optical Sampling
Module
80C11 Optical Sampling
Module
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...