Adjustment Procedures
3- 6
DSA8200 Digital Serial Analyzer and Modules Service Manual
1.
Enter the following GPIB command to turn the instrument’s cal constant
protection off:
SYST:PROT OFF
2.
Set the calibration value to default by sending the following GPIB com-
mand:
CALCOMP:DOUBLE “Internal10MHzRefFreq”,
10e6
3.
Wait 8 seconds then click Run/Stop (button should be green).
4.
Measurement menu, click Show Stats.
5.
Record
target value = (Mean value +1
N
s)
; where
Mean
is from the
Statistics window.
6.
Horizontal menu, Timebase, Position, enter
target value
recorded in step 5.
7.
Click Run/Stop (button should be gray).
8.
Measurement menu, click Clear Stats.
9.
Click Run/Stop (button should be green), wait 8 seconds then click Run/Stop
again (button should be gray).
10.
Using the present
Mean
value from the Statistics window, and the
target value
calculated in step 5,. Calculate the error and the new cal
constant:
Error = (mean -- target value) / 1
N
s
New Cal Const = Error * 10e6 + 10e6
11.
Input the new calibration constant into the instrument by sending the
following GPIB command:
CALCOMP:DOUBLE “Internal10MHzRefFreq”,
New Cal Const
where
New Cal Const
is from Step 10.
12.
Wait 8 seconds then save the mainframe factory calibration constants by
entering the following GPIB commands:
CAL:SAVE:FACT:MAI
13.
Update the mainframe calibration time/date/temperature stamp by entering
the following GPIB command:
CAL:UPDATEINFO:MAI
14.
Enter the following GPIB command to turn the instrument’s cal constant
protection on:
SYST:PROT ON
End of Procedure
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...