Removal and Installation Procedures
DSA8200 Digital Serial Analyzer and Modules Service Manual
4--63
Thermal cover
Holding clips
T--15 Torxdrive
spacer post
Screw
Figure 4--38: Thermal cover removal
j.
Disconnect J6 ribbon cable connector from the large module circuit
board.
k.
Disconnect J5 coaxial cable connector from the large module circuit
board.
l.
Disconnect the three coaxial cables connected to the front panel;
INTERNAL CLOCK OUTPUT, DC CALIBRATION OUTPUT, AND
EXTERNAL 10 MHz REFERENCE INPUT.
m.
Disconnect the TRIGGER PROBE POWER ribbon connector from J25
on the Electrical front end board.
n.
Slide the Acquisition assembly back toward the rear of the instrument
then lift up and out of the instrument.
4.
Reinstallation:
Do, in reverse order, substeps g through n.
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...