Theory of Operation
DSA8200 Digital Serial Analyzer and Modules Service Manual
2- 13
80A01 Trigger Prescale Preamplifier Module
The 80A01 module is designed to increase the sensitivity of the prescale trigger
input of the DSA8200 to
≤
200 mV
pk-pk
.
The major function block of the module is a high sensitivity, high gain RF
amplifier. The input and output to this amplifier are routed to two identical
SMA, female connectors, labelled Input and Output at the module front panel.
The module receives power from the main instrument through a single connector
at the rear of the module. The power LED indicates the module is receiving
power through the interface connector.
For major functional circuit blocks refer to Figure 5--3 on page 5--4.
80A02 EOS/ESD Protection Module
The 80A02 EOS/ESD (Electrical Over Stress/Electro-Static Discharge)
protection module works with any DSA8200 instrument and provides static
electricity damage protection to vulnerable sampling head input stages and/or
other sensitive elements.
The 80A02 EOS/ESD module has a 26 GHz bandwidth, making it possible to
provide static protection to a sensitive single input channel of a sampling
oscilloscope with very minimal speed degradation.
The 80A02 EOS/ESD module is designed to work with either the Tektronix
P8018 probe for manual test station static protection as well as automated test
stations.
For major functional circuit blocks refer to Figure 5--4 on page 5--5.
80A05 Electrical Clock Recovery Module
The electrical clock recovery module is capable of performing clock recovery on
the input signal (signal input must meet data rate and format requirements), and
provides this signal as a trigger source to the DSA8200.
Front panel connectors provide a replica of the recovered clock signal.
The module uses one of two separate clock recovery circuits dependant on the
specified data rate. The single-ended or complementary input signals are split
with a 1:2 divider and routed to the two circuits.
The low bandwidth circuit recovers clock and data from input data in the
50 Mb/s to 2.7 Gb/s range. The recovered clock from this circuit is routed
directly to the front panel and internal trigger.
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...