Troubleshooting
4--90
DSA8200 Digital Serial Analyzer and Modules Service Manual
3.
Enter the following GPIB commands, if necessary install a new instrument’s
serial number:
SYST:PROT OFF
SYST:PROP:MAI:SER <serial number string>, e.g. “B010100”
SYST:PROT ON
4.
These commands take effect immediately and can be verified by viewing the
mainframe properties: Utilities>System Properties.