Removal and Installation Procedures
4--58
DSA8200 Digital Serial Analyzer and Modules Service Manual
1.
Locate module to be removed:
Locate the Fan assembly in the locator
diagram
Internal Modules
, Figure 4--8, page 4--20. Additional modules to be
removed:
1
Trim (all)
1
Bottom cover
1
Left and Right covers
2.
Orient the instrument:
Set the instrument so its bottom is down on the work
surface and its left side is facing you.
3.
Remove the fan assembly:
See Figure 4--35, page 4--59.
CAUTION.
Take care when handling the fan assembly, the fan blades are brittle
and can be easily damaged.
a.
Disconnect the fan from processor/display board:
Disconnect the two fan
power cables J130 and J170 located on the processor board.
b.
Remove the two T-15 Torxdrive screws securing the fan assembly to the
top main chassis.
c.
Lift the fan assembly up and out from the chassis.
4.
Reinstallation:
Do in reverse steps a through c to reinstall the fan assembly.
Fan Assembly Removal
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...