Removal and Installation Procedures
4--78
DSA8200 Digital Serial Analyzer and Modules Service Manual
1.
Orient the module:
Set the module so that its left side is down on the work
surface and its right side is facing you.
CAUTION.
Static discharge can damage any semiconductor in the instrument or
sampling module. To prevent damage from static discharge wear wrist and foot
grounding straps while handling the module chassis. Service should be per-
formed in a static-free work station by a qualified service technician. Minimize
handling of static-sensitive components.
2.
Remove the front panel:
See Figure 4--48, page 4--78.
a.
Remove two coaxial terminators on the front panel connectors.
b.
Remove two Phillips screws that secure the front panel to the cover.
Remove the front panel.
Chassis
Cover
Front panel
Label
Antistatic label
Shaft bracket
Lock-down shaft
Terminators
Figure 4--48: 80A00 and 82A00 series parts removal (80A01 shown)
80A00 and 82A00 Series
Modules
Summary of Contents for DSA8200 Series
Page 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Page 2: ......
Page 6: ......
Page 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 23: ...Operating Information ...
Page 24: ......
Page 33: ...Theory of Operation ...
Page 34: ......
Page 49: ...Adjustment Procedures ...
Page 50: ......
Page 57: ...Maintenance ...
Page 58: ......
Page 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Page 151: ...Diagrams ...
Page 152: ......
Page 161: ...Replaceable Parts ...
Page 162: ......
Page 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...