Setting Jitter/Wander Parameters
CTS850 SDH/PDH Test Set User Manual
3–167
7. To calibrate a 0 dB baseline:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS
JITTER TESTS
Action
Calibrate
8. Press the START/STOP button to run the 0 dB baseline
calibration routine. This takes about 3 minutes.
9. To run the transfer jitter test after the calibration is complete,
press Edit Setup:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS
JITTER TESTS
Action
Jitter Transfer
10. Press the START/STOP button to run the jitter transfer test. The
results are displayed on screen in numerical and graphical form.
This takes abouts 4 minutes for 7 samples. Press View Data to see
the data and whether or not the test passed. Press Edit Setup to
return to Main Jitter Test menu.