Basic Test Procedures
3–12
CTS850 SDH/PDH Test Set User Manual
3. Set up the downstream CTS850 as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS
RECALL
INSTRUMENT
SETUPS
DEFAULT
FACTORY
SETTINGS
Recall Setup
RECEIVE
RECEIVE
SETTINGS
Receive Rate
STM-1E
RESULTS
SDH RESULTS
none
Defects
4. Press the START/STOP buttons on both the upstream and
downstream CTS 850 units so that the START/STOP lights are
on.
5. Verify that RS B1 BIP errors are counted in the SDH RESULTS
display of the upstream CTS850.
6. Verify there are no errors counted in the SDH RESULTS display
of the downstream CTS850.
7. Press the START/STOP buttons on both the upstream and
downstream CTS850s so that the START/STOP lights are off.
8. To check the LTE response to an MS AIS alarm, change the
upstream CTS850 as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TRANSMIT
Defects &
Anomalies
Error Rate set to
None
Transmit Alarm
set to
MS AIS
RESULTS
SDH RESULTS
none
SDH Defects