Table 4-7: Calibration error
Error code
Error log message
Possible defect
on board
Note
200XP0001
Analog Level Calibration CHX positive (negative) 5db attenuator
failed. expected 0.56234, actual xxxxx
OUT10G
OUT10G: K610, K710
200XP0002
Analog Level Calibration CHX positive (negative) 5db attenuator
failed. expected 0.56234, actual xxxxx
OUT10G
OUT10G: K610, K710
200XP0003
Analog Level Calibration CHX positive (negative) 10db attenuator
failed. expected 0.31627, actual xxxxx
OUT10G
OUT10G: K620, K720
200XP0004
Analog Level Calibration CHX positive (negative) 20db attenuator
failed. expected 0.100000, actual xxxxx
OUT10G
OUT10G: K630, K730
200XP0005
Analog Level Calibration CHX positive (negative) 50mhz filter
failed. expected 0.992, actual xxxxx
OUT10G
OUT10G: K660, K760
200XP0006
Analog Level Calibration CHX positive (negative) 200mhz filter
failed. expected 0.998, actual xxxxx
OUT10G
OUT10G: K670, K770
200XP0007
Analog Level Calibration CHX positive (negative) amplitude cal
failed. expected 2.07, actual xxxxx
OUT10G
OUT10G: U800
200XP0008
Analog Level Calibration CHX positive (negative) hfd dac
reference failed.
OUT10G
AWG10G: U100A/B, U150A/B
200XP0009
Analog Level Calibration CHX positive (negative) output offset
failed. expected 0.50V(0.0V), actual xxxxx
OUT10G
OUT10G: U900, U910, U920
200XP0010
Analog Level Calibration CHX internal offset 8bit failed. expected
0.0mV, actual xxxxxmV
OUT10G
AWG10G: U100A/B, U150A/B
200XP0011
Analog Level Calibration CHX internal offset 10bit failed. expected
0.0mV, actual xxxxxmV
OUT10G
AWG10G: U100A/B, U150A/B
200XP0012
Analog Level Calibration CHX can't find 8bit outamp offset cross
point. expected 0.0mV, actual pos: xxxxxmV, neg: xxxxxmV
AWG10G
AWG10G: U100A/B, U150A/B
200XP0013
Analog Level Calibration CHX can't find 10bit outamp offset cross
point. expected 0.0mV, actual pos: xxxxxmV, neg: xxxxxmV
AWG10G
AWG10G: U100A/B, U150A/B
200XP0015
Analog Level Calibration CHX 5db attenuator gain balance failed.
200XP0016
Analog Level Calibration CHX 5db attenuator gain balance failed.
200XP0017
Analog Level Calibration CHX 10db attenuator gain balance failed.
200XP0018
Analog Level Calibration CHX 20db attenuator gain balance failed.
200XP0019
Analog Level Calibration CHX filter 50mhz gain balance failed.
200XP0020
Analog Level Calibration CHX filter 200mhz gain balance failed.
200XP0021
Analog Level Calibration CHX amp gain balance failed.
210XM0001
Marker Level Calibration CHX MARKER M offset failed.
OUT10G
OUT10G: U120, U170, U230,
U240, U330, U340
210XM0002
Marker Level Calibration CHX MARKER M positive offset failed.
OUT10G
OUT10G: U120, U170, U230,
U240, U330, U340
210XM0003
Marker Level Calibration CHX MARKER M negative offset failed.
OUT10G
OUT10G: U120, U170, U230, 240,
U330, U340
210XM0004
Marker Level Calibration CHX MARKER M amplitude failed.
OUT10G
OUT10G: U120, U170, U230,
U240, U330, U340
210XM0005
Marker Level Calibration CHX MARKER M amplitude correction
failed.
OUT10G
OUT10G: U120, U170, U230,
U240, U330, U340
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......