Adjustment Procedures
AWG7000 Series Service Manual
3-13
b.
Modify the sampling oscilloscope settings as follows:
Vertical External Attenuation: 12 dB
Vertical Scale: 200 mV/div
Horizontal Scale: 50 ps/div
Horizontal Timebase Mode: Lock to INT. 10 MHz
Trigger: External Direct Input, Level 50%
Acquisition: Average 64, Stop After Condition: Average Complete
Measurement: R1 to C1 Delay (R1/C1 Reference Level: Absolute
0.0 V)
2.
Click the
Inter-Channel Skew...
button in the Service UI menu to display the
following setup window:
Figure 3-7: Inter-Channel Skew Calibration setup window