Adjustment Procedures
AWG7000 Series Service Manual
3-9
5.
Set
TEK0015 Data Delay
to 0 ns, and observe the oscilloscope screen. If a
sine waveform of approximately 102.4 ns period is output normally, increase
the
TEK0015 Data Delay
setting by 30 ps using the up/down button or the
front-panel rotary knob. You can also enter the value directly using the
keyboard.
6.
Increase the
TEK0015 Data Delay
setting by 30 ps, and observe the
oscilloscope screen. If the waveform starts to break at some point, click
Start
of Broken Waveform
. The following shows an example of broken waveform:
7.
Click
End of Broken Waveform
right before the next waveform is output
normally (the last position where the waveform is broken in that range).
8.
Continue the measurements up to 990 ps in the
TEK0015 Data Delay
setting.
A set value of End of Broken Waveform is always larger than that of Start
of Broken Waveform.
You can temporarily restore the settings to re-check the waveform.
Up/down button
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......