Adjustment Procedures
3-16
AWG7000 Series Service Manual
Figure 3-8: Interleave calibration initial hookup
b.
Modify the sampling oscilloscope settings as follows:
Vertical External Attenuation: 12 dB
Vertical Scale: 200 mV/div
Horizontal Scale: 50 ps/div
Horizontal Timebase Mode: Lock to INT. 10 MHz
Trigger: External Direct Input, Level 50%
Acquisition: Sample
Measurement: C1 Rise Time (C1 Reference Level: Relative
20%-80%)
SMA termination
Instrument under test
Sampling oscilloscope
SMA attenuator
SMA attenuator