Troubleshooting
AWG7000 Series Service Manual
4-55
After Repair
After removal and replacement of a module, you must perform the Adjustment and
Performance Verification procedures. You can find the information about these
procedures in the following manuals:
AWG7000 Series Arbitrary Waveform Generators Service Manual
(Tektronix
part number 071-1854-XX)
AWG7000 Series Arbitrary Waveform Generators Specifications and
Performance Verification
(Tektronix part number 071-1853-XX)
300000217
An attempt was made to load a program with an incorrect
format.(PLD130b.Bit)
N/A
300000218
The DCM timeout period has expired.
MIO
MIO: U300
300000219
The LOCKED timeout period has expired.
MIO
MIO: U300
300000220
A PLD131.bit file is old version.
N/A
300000221
A PLD130.bit file is old version.
N/A
310000000
The parameter is incorrect.
N/A
390000001
The calibration data is invalid. The default value is used.
N/A
390000002
*** Warning *** It started with Manufacturing mode.
N/A
390000003
The calibration data is invalid. The default value is used.
N/A
Table 4-8: Hardware error (cont.)
Error code
Error log message
Possible defect
on board
Note
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......