Troubleshooting
AWG7000 Series Service Manual
4-47
131000002
HwDiagClockPll 1GHz PLL Locked
AWG10G, CLK10G AWG10G: U1000
CLK10G: U3200
131000003
HwDiagClockPll 4.9GHz YIG tuning error
AWG10G, CLK10G AWG10G: U1000, U1040
CLK10G: U3200,U3250,Y3100,Y3280
131000004
HwDiagClockPll 4.9GHz PLL Unlocked
AWG10G, CLK10G AWG10G: U1000, U1040
CLK10G: U3200,U3250,Y3100,Y3280
131000005
HwDiagClockPll 10.1GHz YIG tuning error
AWG10G, CLK10G AWG10G: U1000, U1040
CLK10G: U3200,U3250,Y3100,Y3280
131000006
HwDiagClockPll 10.1GHz PLL Unlocked
AWG10G, CLK10G AWG10G: U1000, U1040
CLK10G: U3200,U3250,Y3100,Y3280
140000000
The error occurred in the EEPROM read. 1st=
FST
2nd=
SND
FST: the data that is received from EEPROM at the first read
check
SND: the data that is received from EEPROM at the second
read check, which is implemented after writing 0x55 to
EEPROM
OUT10G
OUT10G-CH1: U500, U550
140000001
Message varies depending on the situation.
OUT10G
OUT10G-CH1: U500, U550
1410P0001
Analog Level Diag CH1 positive (negative) 5db attenuator
failed. expected 0.56234, actual xxxxx
OUT10G
OUT10G: K610, K710
1410P0002
Analog Level Diag CH1 positive (negative) 5db attenuator
failed. expected 0.56234, actual xxxxx
OUT10G
OUT10G: K610, K710
1410P0003
Analog Level Diag CH1 positive (negative) 10db attenuator
failed. expected 0.31627, actual xxxxx
OUT10G
OUT10G: K620, K720
1410P0004
Analog Level Diag CH1 positive (negative) 20db attenuator
failed. expected 0.100000, actual xxxxx
OUT10G
OUT10G: K630, K730
1410P0005
Analog Level Diag CH1 positive (negative) 50mhz filter
failed. expected 0.992, actual xxxxx
OUT10G
OUT10G: K660, K760
1410P0006
Analog Level Diag CH1 positive (negative) 200mhz filter
failed. expected 0.998, actual xxxxx
OUT10G
OUT10G: K670, K770
1410P0007
Analog Level Diag CH1 positive (negative) amplitude cal
failed. expected 2.07, actual xxxxx
OUT10G
OUT10G: U800
1410P0008
Analog Level Diag CH1 positive (negative) hfd dac
reference failed.
AWG10G
AWG10G: U100A, U150A
1410P0009
Analog Level Diag CH1 positive (negative) output offset
failed. expected 0.50V(0.0V), actual xxxxx
AWG10G
OUT10G: U900, U910, U920
141000010
Analog Level Diag CH1 internal offset 8bit failed. expected
0.0mV, actual xxxxxmV
AWG10G
AWG10G: U100A, U150A
141000011
Analog Level Diag CH1 internal offset 10bit failed. expected
0.0mV, actual xxxxxmV
AWG10G
AWG10G: U100A, U150A
141000012
Analog Level Diag CH1 can't find 8bit outamp offset cross
point. expected 0.0mV, actual pos: xxxxxmV, neg: xxxxxmV
AWG10G
AWG10G: U100A, U150A
141000013
Analog Level Diag CH1 can't find 10bit outamp offset cross
point. expected 0.0mV, actual pos: xxxxxmV, neg: xxxxxmV
AWG10G
AWG10G: U100A, U150A
141000015
Analog Level Diag CH1 5db attenuator gain balance failed.
OUT10G
OUT10G: K610, K710
141000016
Analog Level Diag CH1 5db attenuator gain balance failed.
OUT10G
OUT10G: K610, K710
Table 4-6: Diagnostics error (cont.)
Error code
Error log message
Possible defect
on board
Note
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......