Hardware Error
Table 4-8 lists the instrument hardware related errors.
270000002
HwExecYigCal PLL 5GHz offset underflow vtune=xxxx
CLK10G
CLK10G: U3200, U3250, Y3280
270000003
HwExecYigCal PLL 5GHz tuning underflow vtune=xxxx
CLK10G
CLK10G: U3200, U3250, Y3280
270000004
HwExecYigCal PLL 5GHz tuning underflow vtune=xxxx
CLK10G
CLK10G: U3200, U3250, Y3280
270000005
HwExecYigCal PLL 10GHz tuning over 4095 vtune=xxxx
CLK10G
CLK10G: U3200, U3250, Y3280
270000006
HwExecYigCal PLL 10GHz tuning overflow vtune=xxxx
CLK10G
CLK10G: U3200, U3250, Y3280
290000000
Marker Level Calibration has not been completed.
N/A
290000001
Marker cross point calibration result exceeds limit.
OUT10G
OUT10G: U120, U170
290X00000
Marker Level Calibration has not been completed at CHX.
N/A
290XY0001
Marker cross point calibration result exceeds limit at CHX,
MARKER Y.
OUT10G,
AWG10G
OUT10G: U120/170, U470/480,
J100/110/150/170
AWG10G: J110A/111A/115A/
116A/110B/111B/115B/116B
290XY0002
Marker cross point calibration couldn't be converged on -1.0V at
CHX, MARKER Y.
OUT10G,
AWG10G
OUT10G: U120/170, U470/480,
J100/110/150/170
AWG10G: J110A/111A/115A/
116A/110B/111B/115B/116B
Table 4-7: Calibration error (cont.)
Error code
Error log message
Possible defect
on board
Note
Table 4-8: Hardware error
Error code
Error log message
Possible defect
on board
Note
300000000
The system cannot find the device specified.
MIO
MIO: U100, U200, U220,Y230
300000001
The device is not ready.
N/A
300000002
The AWG register access timeout period has expired.
MIO: U200, U300
AWG10G: U1000
MIO-AWG10G Cable
300000003
The time-out was generated while measuring data.
AWG10G
AWG10G: U1000
300000004
The timer for the trigger has stopped.
AWG10G
AWG10G: U1000
300000005
PLL doesn't lock.
AWG10G,
CLK10G
AWG10G: U1000
CLK10G: U3200, U3250,Y3260...
300000006
AwgController DLL doesn't lock.
AWG10G
AWG10G: U1000
300000007
Output board access error Board=
No
No: Board number (Start with 0)
OUT10G
OUT10G: U500
300000008
Not enough storage is available to process.
N/A
300000009
Calibration checksum data has an error.
MIO
MIO: U300,U390
300000010
Calibration version unmatched.
MIO
MIO: U300,U390
300000011
YIG tuning has an error.
CLK10G
CLK10G: U3200, U3250,Y3260...
Summary of Contents for AWG7000 Series
Page 4: ......
Page 10: ...List of Tables vi AWG7000 Series Service Manual...
Page 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Page 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Page 18: ...Preface xiv AWG7000 Series Service Manual...
Page 19: ...Operating Information...
Page 20: ......
Page 23: ...Theory of Operation...
Page 24: ......
Page 31: ...Adjustment Procedures...
Page 32: ......
Page 51: ...Maintenance...
Page 52: ......
Page 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Page 109: ...Replaceable Parts...
Page 110: ......