Figure 9: External Trigger Direct Input jitter test setup
1.
Install the 80E0X module(s) to test starting at the left-most small module
slot as shown in the
fi
gure. Power on the instrument and wait 20 minutes
for warm-up.
2.
Connect the test equipment (See Figure 9.)
3.
Set the synthesized signal generator as follows:
Frequency to
2.5 GHz
.
Output to
1.6 V
p-p
(800 mV
p-p
displayed).
4.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select
C1
from the Waveform Selector menu (left end of controls bar
at bottom of the graticule).
c.
Click
Setup > Mode/Trigger
and select
Direct
for the Trigger Source.
d.
Click the
Vert
tab and set the following:
Scale to
100 mV/div
.
Offset to
0
.
e.
Click the
Horz
tab and set the following:
Scale to
200 ps/division
.
Position to
19 ns
(29 ns if using 80E07B, 80E08B, 80E09B, 80E10B).
Record Length to
2000
.
Horizontal Reference to
0%
.