![Tektronix 80A00 Technical Reference Download Page 182](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946182.webp)
DSA8300 Performance Veri
fi
cation
15.
Set the Vertical Scale to
5 mV/div
.
16.
Click the
Acq
tab and set the following:
Stop After to
Condition
.
Number of Acquisitions to
400
.
17.
Push the
CLEAR DATA
front-panel button, and then push the
RUN/STOP
button. The instrument takes 400 acquisitions and stops. This acquisition
takes a couple of seconds. Note the value in the RMS Jitter measurement
readout.
18.
Repeat step to take several RMS Jitter readings and enter the average value on
the Test Record for the module under test. Typical values are
200 fs or less
.
This measurement does not have a guaranteed speci
fi
cation.
19.
Move the signal source cable to the next module under test. Repeat from setup
step for each 80E00 series modules installed. Check one input per module
(for example, channels 1 and 3 if two modules are installed).
20.
Disconnect the test setup.
End of Procedure
Trigger Delay Jitter, External Triggered Phase Corrected Mode - Clock Eye
Equipment required
82A04 Phase Reference module
0.5 meter precision SMA cable (item 15)
1.0 meter precision SMA cable (item 16)
Power divider (item 26)
Synthesized signal generator (item 31)
Electrical sampling module (80E01, 80E02, 80E03, 80E04,
80E06, 80E07B, 80E08B, 80E09B, and 80E10B)