![Tektronix 80A00 Technical Reference Download Page 171](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946171.webp)
DSA8300 Performance Veri
fi
cation
e.
Click the
Horz
tab and set the following:
Set the Scale to
200 ps/div
.
Set Position to
19 ns
(29 ns if using 80E07B, 80E08B, 80E09B, or
80E10B).
Set Horizontal Reference to
0%
.
f.
Click the Acq tab and set Acquisition Mode to
Average
.
Test
Follow this procedure to make the time interval accuracy check:
1.
Click the
Meas
tab and set the following:
2.
Select
Meas 1
.
3.
Click the
Source1
button.
4.
In the Source tab, make sure the channel under test is selected and click the
Pulse
Signal Type button.
5.
Click
Select Meas > Pulse – Amplitude > Amplitude
.
6.
Click
Meas 2
.
7.
Click the
Source1
button.
8.
In the Source tab, make sure the channel under test is selected and click the
Pulse
Signal Type button.
9.
Click
Select Meas > Pulse – Timing > Period
.
10.
Adjust the synthesizer signal generator level until the amplitude measurement
readout on the instrument screen is
400 mV
.
11.
Verify that the period measurement readout value is
1 ns ±9 ps
. Record the
value in the
Time Interval Accuracy, Trigger Direct Input connector
test
record
fi
eld.
12.
Click the
HORZ
tab, set the Horizontal Position
fi
eld to
100 ns
, and verify
that the period measurement readout value is
1 ns ±9 ps
. Record the value in
the
Time Interval Accuracy, Trigger Direct Input connector
test record
fi
eld.
13.
Set the Horizontal Position
fi
eld to
1
μ
s
and verify that the period measurement
readout value is
1 ns ±9 ps
. Record the value in the
Time Interval Accuracy,
Trigger Direct Input connector
test record
fi
eld.
14.
Disconnect the 0.5 m cable from the TRIGGER DIRECT INPUT connector
and connect it to the
CLOCK INPUT/PRESCALE TRIGGER
input.
15.
Click the
Mode/Trigger
tab and click the
Clock
trigger source button.
16.
Click the
Eye
button (Scope Mode (Clock Trigger Source) area).
DSA8300 Performance Veri
fi
cation
153