![Tektronix 80A00 Technical Reference Download Page 186](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946186.webp)
DSA8300 Performance Veri
fi
cation
Setup
Figure 12: External Clock/Prescale trigger delay jitter test setup
1.
Install the 80E0X module(s) to test into the left-most small module slots.
Power on the instrument and wait 20 minutes for warm-up.
2.
Connect the test equipment (See Figure 12.)
3.
Set the synthesized signal generator as follows:
Frequency to
2.5 GHz
.
Output to
1.6 V
p-p
(800 mV
p-p
displayed).
4.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select the input channel of the module under test from the Waveform
Selector menu (left end of controls bar at bottom of the graticule).
c.
Select
Setup > Mode/Trigger
tab and set the following:
Trigger Source to
Clock
.
Click the
Eye
button (Scope Mode area).
d.
Push the
Clear Data
front panel button.
168
DSA8300 Performance Veri
fi
cation
Summary of Contents for 80A00
Page 2: ......
Page 6: ......
Page 16: ...Important safety information x DSA8300 Performance Verification ...