Electrical Sampling Modules
Setup
Install the test setup to measure reference signals:
Figure 19: Measure reference signals
NOTE.
Use a torque wrench while connecting the adapters, power sensor in this
setup. Use 8 ±0.3 inch pounds of torque.
1.
Connect the cable to the Synthesized signal generator. Connect the remaining
end of the cable to the 10 dB attenuator.
2.
Connect the other end of the 10 dB attenuator to a 2.4 mm female-to-female
adapter.
3.
Connect the other end of the 2.4 mm female-to-female adapter to the power
sensor.
4.
Connect the power sensor to the power meter.
5.
Set the synthesized signal generator to
+3 dBm
.
6.
Set the signal frequency and the power meter frequency to each of the
frequencies listed in the
Frequency
column of the following table.
NOTE.
Only perform checks up to the maximum bandwidth of the module under
test.
188
DSA8300 Performance Veri
fi
cation