Electrical Sampling Modules
5.
Connect one 50
Ω
precision coaxial cable to the output of a DC power supply
through a dual-banana connector. Connect the remaining 50
Ω
precision
coaxial cable to a digital multimeter as shown in the following
fi
gure.
6.
Set up the main instrument as follows:
a.
Select
C1
from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
b.
In the Mode/Trigger dialog box, set Trigger Source to
TDR
.
c.
In the Acq Setup dialog box:
Set Acquisition Mode to
Average
and
32 samples
.
Under Stop After, select the
Condition
box, and select
Average
Complete
from the pull-down menu.
d.
In the Horz Setup dialog box:
Set Scale to
2 ns/div
.
Set Position to
0
(this will automatically change to a minimum
non-zero value).
Set Horizontal Reference to
0%
.
e.
In the Meas Setup dialog box:
Select
Meas 1
and make sure the
Source1
button is selected.
Make sure the channel under test is selected in the Source tab, click
the
Pulse
Signal Type button, and then choose
Select Meas > Pulse -
Amplitude > Mean
.
178
DSA8300 Performance Veri
fi
cation