![Tektronix 80A00 Technical Reference Download Page 172](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946172.webp)
DSA8300 Performance Veri
fi
cation
17.
Click the
Clear Data
button (application toolbar; looks like a red and white
eraser).
18.
Click the
Horz
tab and set the Position
fi
eld to
19 ns
(29 ns if using 80E07B,
80E08B, 80E09B, or 80E10B).
19.
Repeat steps 11 through 13. Record the value in the
Time Interval Accuracy,
Clock Input/Prescale Trigger connector; Eye mode
test record
fi
eld.
20.
Click the
Mode/Trigger
tab and click the
Other
button (Scope Mode (Clock
Trigger Source) area).
21.
Click the
Horz
tab and verify/set the Position
fi
eld to
19 ns
(29 ns if using
80E07B, 80E08B, 80E09B, or 80E10B).
22.
Repeat steps 11 through 13. Record the value in the
Time Interval Accuracy,
Clock Input/Prescale Trigger connector; Other mode
test record
fi
eld.
End of Procedure
External Trigger Direct Level Accuracy
Equipment required
0.5 meter precision SMA cable (item 15)
1.0 meter precision SMA cable (item 16)
Power divider (item 26)
Synthesized signal generator (item 31)
Electrical sampling module (80E01, 80E02, 80E03, 80E04,
80E06, 80E07B, 80E08B, 80E09B, or 80E10B)
Prerequisites
As listed under
Prerequisites
Setup
Figure 7: External direct trigger test setup