![Tektronix 80A00 Technical Reference Download Page 176](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946176.webp)
DSA8300 Performance Veri
fi
cation
Test
Follow this procedure to make the direct trigger sensitivity check:
1.
Click the
Meas
tab.
2.
Select
Meas 1
.
3.
Click the
Source1
button.
4.
Make sure the channel under test is selected in the
Source
tab
5.
Click the
Pulse
button (Signal Type area).
6.
Click
Select Meas > Pulse – Amplitude > Amplitude
.
7.
Click
Meas 2
.
8.
Click the
Source1
button.
9.
In the Source tab, make sure the channel under test is selected and click the
Pulse
Signal Type button.
10.
Click
Select Meas > Pulse – Timing > Frequency
.
11.
Adjust the synthesizer signal generator level until the amplitude measurement
readout on the instrument shows
100 mV
.
12.
Set the signal generator frequency to
100 MHz
.
13.
Adjust the Horizontal
SCALE
knob to display at least two full periods.
14.
Verify that the frequency readout value is within
10%
of the true value.
15.
Set the signal generator frequency to each of the following values, adjust the
horizontal scale to the corresponding seconds/div setting, and repeat steps 13
and 14 :
800 MHz at 500 ps/div
1 GHz at 500 ps/div
2 GHz at 200 ps/div
3 GHz at 100 ps/div
End of Procedure
External Trigger Direct Input Connector Delay Jitter
Equipment required
0.5 meter precision SMA cable (item 15)
1.0 meter precision SMA cable (item 16)
Power divider (item 26)
Synthesized signal generator (item 31)
Electrical sampling module (80E01, 80E02, 80E03, 80E04,
80E06, 80E07B, 80E08B, 80E09B, or 80E10B)
Prerequisites
As listed under
Prerequisites