80A02 EOS/ESD Protection Module
Test
Follow this procedure to make the control threshold check:
1.
Check the auto control operation of the 80A02 module:
a.
Check that the relay in the 80A02 module does not activate (no click
sound) and that the ENGAGED LED does not light while adjusting the
DC supply from
0
to
+300 mV
.
b.
Check that the relay in the 80A02 module activates (click sound) and that
the ENGAGED LED lights while adjusting the DC supply from
+500 mV
(+0.5 V)
to
+5 V
.
c.
Disconnect the voltage supply.
2.
Remove the mini phone plug.
End of Procedure
252
DSA8300 Performance Veri
fi
cation