![Tektronix 80A00 Technical Reference Download Page 272](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946272.webp)
80A05 Electrical Clock Recovery Module
Setup
Figure 39: Clock recovery sensitivity range and recovered clock timing jitter setup
1.
Install the 80A05 Clock Recovery Module into Channel 1 of the instrument.
2.
Install the 80E02 Sampling Module into Channels 3 and 4 of the instrument.
3.
Connect a less than 2 ns delay, 50
Ω
coaxial cable between the 1/32 CLK
signal output of the pattern generator and the TRIGGER DIRECT INPUT
connector on the instrument.
4.
Install 2 10X (20 dB) SMA electrical attenuators (in series, creating 100X
attenuation) on the ELECTRICAL INPUT IN+ of the 80A05 Module.
5.
Connect a less than 2 ns delay, 50
Ω
coaxial cable between the DATA +
OUTPUT of the pattern generator and the ELECTRICAL INPUT IN+ of the
80A05.
254
DSA8300 Performance Veri
fi
cation
Summary of Contents for 80A00