![Tektronix 80A00 Technical Reference Download Page 189](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946189.webp)
DSA8300 Performance Veri
fi
cation
1.
Install the 80E0X module(s) to test into the left-most small module slots.
Power on the instrument and wait 20 minutes for warm-up.
2.
Connect the test equipment (See Figure 13.)
3.
Set the synthesized signal generator as follows:
Frequency to
2.0 GHz
(500 ps period).
Output to
400 mV
p-p
(200 mV
p-p
displayed).
4.
Set the DSA8300 as follows:
a.
Push the
Default Setup
front-panel button and click
Yes
.
b.
Select the input channel of the module under test from the Waveform
Selector menu (left end of controls bar at bottom of the graticule).
c.
Select
Setup > Mode/Trigger
and set the following:
Trigger Source to
Clock
.
Click the
Eye
button (Scope Mode area).
d.
Click the
Vert
tab and set the following:
Scale to
50 mV/div
.
Offset to
0
.
e.
Click the
Horz
tab and set the following:
Scale to
200 ps/div
.
Horizontal Reference to
0%
.
Position to
20 ns
(set to
29 ns
for 80E07B, 80E08B, 80E09B, and
80E10B).
Test
Follow this procedure to make the External Clock/Prescale Trigger Sensitivity
check:
1.
Click the
Meas
tab.
2.
Select
Meas 1
.
3.
Click the
Source1
button.
4.
Make sure the channel under test is selected in the
Source
tab
5.
Click the
Pulse
button (Signal Type area).
6.
Click
Select Meas > Pulse – Amplitude > Amplitude
.
7.
Select
Meas 2
.
8.
Click the
Source1
button.
DSA8300 Performance Veri
fi
cation
171
Summary of Contents for 80A00