![Tektronix 80A00 Technical Reference Download Page 211](http://html1.mh-extra.com/html/tektronix/80a00/80a00_technical-reference_1077946211.webp)
Electrical Sampling Modules
7.
Record amplitude reading readout and use the value as AMPLITUDE (Volts)
in the step that follows.
8.
Convert Volts to dBm as follows:
Vp=(AMPLITUDE/2)
V
RMS
=Vp/SQRT (2)
P1=V
RMS
*V
RMS
/50
dBm=10*log(P1/0.001)
9.
Record dBm in the
Measured level
column of the table. (See Table 8.)
10.
Repeat steps 2 through 9 for all frequencies listed in the
Frequency
column
of the table. (See Table 8.)
11.
Calculate the DUT response as the difference between the
Measured level
column and the
Reference RF level
column in the table (See Table 8.) as:
(Measured level) - (Reference RF level)
Record the calculation results in the
DUT Response
column of the table. (See
12.
Calculate all frequencies listed in the
Frequency
column of the table. (See
13.
Find the maximum and minimum power points in the
DUT Response
column
of the table. (See Table 8.)
14.
The maximum power minus the power at 50 MHz should be less than +3.0 dB.
The minimum power minus the power at 50 MHz should be greater than
–3.0 dB. The speci
fi
cation is ±3.0 dB referenced to 50 MHz.
15.
Repeat the Setup and Test sections for all bandwidth settings that are valid for
the module under test.
16.
Repeat the Setup and Test sections for channel 2.
17.
Disconnect the test setup.
End of Procedure