Operation and
maintenance manual
MU_029_EN
07/11/18
EN
Rev.0
Page 30 of 35
Immunity test
Test level to
EN/IEC 60601-1-2
Conformity
level
Electromagnetic environment -
directives
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Veff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
3 Veff
3 V/m
Portable
and
mobile
RF
communications equipment should
be used no closer to any part of the
Products, included cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended
separation
distance:
d = 1,2
√
P
150 KHz to 80 MHz
d = 1,2
√
P
80 MHz to 800 MHz
d = 2,3
√
P
800 MHz to 2.7 GHz
where
P
is the maximum output
power rating of the transmitter in
watts (W), according to the
transmitter manufacture and
d
is
the
recommended
separation
distance in meters (m).
Field
strengths
from
fixed
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance leave
in each frequency range.
Interference may occur
in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects an people.
Summary of Contents for ORION TECH
Page 2: ......
Page 37: ...Manuale d uso e manutenzione MU_029_IT 07 11 18 IT Rev 0 Pag 35 di 35 Note...
Page 38: ......
Page 73: ...Operation and maintenance manual MU_029_EN 07 11 18 EN Rev 0 Page 35 of 35 Notes...
Page 74: ......
Page 109: ...Manual de uso y mantenimiento MU_029_ES 07 11 18 ES Rev 0 P g 35 de 35 Notas...
Page 110: ......
Page 145: ...Notice d utilisation et d entretien MU_029_FR 07 11 18 FR R v 0 Page 35 de 35 Notes...
Page 146: ......
Page 147: ...Drawing code Rev Data 193 1 3 2 4 5 6 9 8 7...
Page 148: ...Drawing code Rev Data 19 11 2018 220 1 2 3 4 5 6 7 9 10 8...
Page 149: ...Drawing code Rev Data 195 1 2 3 4 5 6 7 8 1 08 01 18...
Page 150: ...Drawing code Rev Data 198 3 4 1 2 2 2 1 2...
Page 151: ...Drawing code Rev Data 199 1 2 5 6 7 3 4...
Page 152: ...Drawing code Rev Data 200 2 3 4 1 7 5 6...
Page 153: ...E Drawing code Rev Data 201 1 2 4 3 3 8 7 9 8 5 6...
Page 154: ......
Page 155: ......
Page 156: ......