Operation and
maintenance manual
MU_029_EN
07/11/18
EN
Rev.0
Page 19 of 35
The handpieces can withstand about 200 steam sterilization cycles in
accordance with the following parameters:
-
steam sterilization at 121°C 1.3 bar for 25 to 30 minutes
-
steam sterilization at 134°C 2.3 bar for 4 minutes
Do not exceed a sterilization temperature of 134°C.
Strictly keep to the ISO 17665-1 standard
When placing in the autoclave, make sure the open side of the
handpieces is turned downwards. The handpieces must be free and not
burdened by other material being sterilized.
Damaged handpieces must no longer be used.
7 Adjustment and maintenance
7.1 Installation of swinging arm
See drawing 198
The Product is sold already balanced and does not require further
adjustment. In the event of the spring swinging arm becoming stiff or
loose over time, mechanical intervention is possible by regulating the
compression of the internal spring.
Manually move the cover (1) forwards. Fit the adjustment pin (2) in the
holes of the ring nut and turn in the direction indicated by the arrows to
increase/decrease the load on the spring.
If the swinging arm drops, this means the elastic force of the spring is
insufficient:
- turn the lever upwards and release the spring.
If the swinging arm lifts up, this means the elastic force of the spring is
too high:
- turn the lever upwards and release the spring.
At the end of adjustment, manually reposition the cover (1) in its original
position.
The swinging movement of the arm can also be adjusted upwards.
The Product is sold with maximum set swinging movement. If the
swinging movement is to be reduced upwards, manually move the cover
(1) forwards and insert the adjustment pin (2) in the second ring nut. By
turning the pin downwards, the swinging movement can be reduced until
it is in horizontal position.
The swinging movement downwards cannot instead be changed.
Summary of Contents for ORION TECH
Page 2: ......
Page 37: ...Manuale d uso e manutenzione MU_029_IT 07 11 18 IT Rev 0 Pag 35 di 35 Note...
Page 38: ......
Page 73: ...Operation and maintenance manual MU_029_EN 07 11 18 EN Rev 0 Page 35 of 35 Notes...
Page 74: ......
Page 109: ...Manual de uso y mantenimiento MU_029_ES 07 11 18 ES Rev 0 P g 35 de 35 Notas...
Page 110: ......
Page 145: ...Notice d utilisation et d entretien MU_029_FR 07 11 18 FR R v 0 Page 35 de 35 Notes...
Page 146: ......
Page 147: ...Drawing code Rev Data 193 1 3 2 4 5 6 9 8 7...
Page 148: ...Drawing code Rev Data 19 11 2018 220 1 2 3 4 5 6 7 9 10 8...
Page 149: ...Drawing code Rev Data 195 1 2 3 4 5 6 7 8 1 08 01 18...
Page 150: ...Drawing code Rev Data 198 3 4 1 2 2 2 1 2...
Page 151: ...Drawing code Rev Data 199 1 2 5 6 7 3 4...
Page 152: ...Drawing code Rev Data 200 2 3 4 1 7 5 6...
Page 153: ...E Drawing code Rev Data 201 1 2 4 3 3 8 7 9 8 5 6...
Page 154: ......
Page 155: ......
Page 156: ......