123
3.
Select either [1-point Adjustment] or [2-point Adjustment].
4.
Please enter the actual pre-adjusted value(s) in the [Current Reading] column and
the post-adjusted reference value(s) in the [Post-adjusted Reading] column.
5.
Click on the [Send Settings] Button to send the adjustment equation to the
selected Data Logger.
[Rough Guidelines for Adjustment]
1 point Adjustment: Use when measuring in a range of
Ⱦ
20ºC.
This adjustment method changes only the offset (b) where the slope (a) is considered 1.
For example, the RTR-52Pt is measuring 10ºC, but the standard says the actual measurement
should be 10.2ºC. Enter 10 in the Current Reading box and 10.2 in the Post - adjusted Reading
box
The conversion equation will be Y=X+0.2 and for all measurements an adjustment to the offset of
+0.2 will be made.